Turk Kardiyol Dern Ars. 2013; 41(6): 522-525 | DOI: 10.5543/tkda.2013.44845
Safe right bundle branch block pattern during permanent right ventricular pacing: a case report
Abdelkarim Errahmouni, Jean Christophe Charniot, Jean Yves ArtigouDepartment of Cardiology, Avicenne Hospital, (Assistance Publique Des Hôpitaux De Paris), Bobigny, France
We present the case of a 72-year-old male who underwent a permanent transvenous DDD pacemaker implantation for symptomatic complete heart block. The postoperative surface electrocardiogram (ECG) revealed a right bundle branch block (RBBB) pacing pattern with transition in lead V3 and left bundle branch block (LBBB) in D1, suggesting a malpositioned ventricular lead in the left heart. Malposition of a lead was ruled out by chest X-ray and echocardiogram confirming apical right ventricular pacing. After a literature review, a surface ECG maneuver, by placing leads V1 and V2 one intercostal space lower, restored the usual patterns of paced LBBB pattern. We discuss the literature concerning safe-paced RBBB.
Keywords: Bundle-branch block/etiology, cardiac pacing, artificial, echocardiography; heart block/therapy.