ISSN 1016-5169 | E-ISSN 1308-4488
Archives of the Turkish Society of Cardiology
Pacemakers do not always tell the truth: diagnosis of ventricular tachycardia for supraventricular tachycardia on pacemaker telemetry [Turk Kardiyol Dern Ars]
Turk Kardiyol Dern Ars. 2011; 39(8): 683-685 | DOI: 10.5543/tkda.2011.01687

Pacemakers do not always tell the truth: diagnosis of ventricular tachycardia for supraventricular tachycardia on pacemaker telemetry

Fatih Bayrak, Carlo de Asmundis, Chierchia Gian-Battista, Pedro Brugada
Uz Brussels-vub, Heart Rhythm Management Center, Brussels, Belgium

Pacemaker telemetry reflects algorithm-dependent interpretations about the cardiac rhythm. For these data to be useful, it is necessary to recognize the limitations of each algorithm and its interactions with pacing and sensing parameters. We report on a case in which pacemaker Holter recording showed several episodes of nonsustained supraventricular tachycardia which were erroneously classified as ventricular tachycardia.

Keywords: Pacemaker, arrhythmia detection algorithm

How to cite this article
Fatih Bayrak, Carlo de Asmundis, Chierchia Gian-Battista, Pedro Brugada. Pacemakers do not always tell the truth: diagnosis of ventricular tachycardia for supraventricular tachycardia on pacemaker telemetry. Turk Kardiyol Dern Ars. 2011; 39(8): 683-685

Corresponding Author: Fatih Bayrak, Türkiye