Dislodgement of a sirolimus-eluting stent in the circumflex artery and its successful deployment with a small-balloon technique [Turk Kardiyol Dern Ars]
Turk Kardiyol Dern Ars. 2011; 39(5): 418-421 | DOI: 10.5543/tkda.2011.01248

Dislodgement of a sirolimus-eluting stent in the circumflex artery and its successful deployment with a small-balloon technique

Tunay Şentürk, Bülent Özdemir, Dilek Yeşilbursa, Osman Akın Serdar
Department of Cardiology, Uludag University School of Medicine, Bursa, Turkey

Coronary stent dislodgement or embolization before deployment is a rare but serious complication in interventional cardiology. A 60-year-old male presented with unstable angina five years after coronary artery bypass surgery. There was a stenosis (70%) in the obtuse marginal branch of the circumflex artery. During percutaneous coronary intervention, a sirolimus-eluting stent was stripped from its balloon mainly because of significant proximal angulation and incarcerated within the proximal circumflex artery. A smaller balloon dilatation catheter was advanced and pushed through the inside of the slipped stent. Using this technique, the stent could be advanced into the lesion and was successfully deployed. To our knowledge, this is the first case report on sirolimus-eluting stent dislodgement.

Keywords: Angioplasty, balloon, coronary/instrumentation, equipment failure; stents

How to cite this article
Tunay Şentürk, Bülent Özdemir, Dilek Yeşilbursa, Osman Akın Serdar. Dislodgement of a sirolimus-eluting stent in the circumflex artery and its successful deployment with a small-balloon technique. Turk Kardiyol Dern Ars. 2011; 39(5): 418-421

Corresponding Author: Tunay Şentürk, Türkiye
Manuscript Language: English


Journal Metrics

Journal Citation Indicator: 0.18
CiteScore: 1.1
Source Normalized Impact
per Paper:
0.22
SCImago Journal Rank: 0.348

Quick Search



Copyright © 2024 Archives of the Turkish Society of Cardiology



Kare Publishing is a subsidiary of Kare Media.