Between September 1990 and July 1994, 26 transcatheter occlusion of persistently patent ductus arteriosus was attempted in 25 patients using Rashkind umbrella occluder. The patients ages ranged from 2 years to 26 years (median 5.04 years) and patients' weight ranged from 11.5 kg to 52 kg (median 17 kg). In 12 patients a 12 mm and in 12 a 17 mm diameter device were successfully placed inthe ductus (internal diameter average 3.86±1.75 mm, range 2.25 to 9 mm). One procedure was abandoned because of fluoroscopy insufficiency. Significant residual shunt with continuous murmur was present on the postocclusion aortography in 3 of 24 patients. Complete closure of the ductus with implantation of a second occluder device was achieved in one of these 3 patients. Twelve patients had total occlusion and the remaining 9 patients showed minimal residual leaks on postocclusion aortography. No device embolisation occurred. Catheter occlusion of the patent ductus arteriosus using the Rashkind umbrella appears to be a safe and effective method of nonsurgical management.
Keywords: The patent ductus arteriosus, transcatheter occlusionCopyright © 2024 Archives of the Turkish Society of Cardiology